If you are having trouble logging into your existing account, please reset your password. If the problem persists, please let us know.
Free Shipping on Orders $175+. Read more.

Hanna HI7061M Cleaning Solution General Purpose

Sku: HI7061M

More Views

Hanna HI7061M Cleaning Solution General Purpose

Sku: HI7061M

$11.00

Availability: Factory Direct - Usually Ships Out Within 3-4 Days

Be the first to review this product

Hanna HI7061M Cleaning Solution General Purpose

You will earn 11 Points for buying this product
OR
Description

Details

Hanna Instruments HI7061M Cleaning Solution General Purpose

HI7061M is a general purpose cleaning solution that can be used for a variety of applications. Electrodes can become dirty from use and will produce inaccurate results even as they read correctly in a pH buffer. Hanna’s cleaning solutions eliminate impurities and residues that are left on electrode surfaces when immersed in samples during measurement or when stored incorrectly. Hanna suggests cleaning the bulb and junction of your electrode on a regular basis to ensure that the probe is always clean and to prevent any clogging of the junction.

Specially formulated for cleaning electrodes

Air tight seal to ensure quality

Expiration date and lot number

Details

The most common cause for pH measurement inaccuracies is an unclean or improperly cleaned electrode. This is very important to note, because during calibration, the instrument assumes that the electrode is clean and that the standardization curve created during the calibration process will remain a valid reference until the next calibration. pH meters on the market today will allow an offset voltage of approximately ±60 mV. The deviation from 0 mV is not unusual but ideally should be no greater than ±30 mV. The calibration process compensates for the change in offset voltage. If the mV offset continues to deviate and the meter is calibrated with a dirty electrode then inaccurate readings will result.

Cleaning is a fast and effective routine that should be performed on a regular basis as a preventative measure against using a dirty electrode and to ensure that the junction is not clogged. By simply immersing the electrode in the HI7061M cleaning solution for 15 to 20 minutes will dissolve any mineral deposits or other coatings.  

    Air tight bottle with tamper-proof seal
  • Ensures quality and freshness of solution
    Each bottle marked with lot number and expiration date.
  • Hanna cleaning solutions are specially formulated to have an expiration of 5 years from manufacture for an unopened bottle.

Specifications

Weight .55

Application General Purpose
Package bottles
Size 230 mL
Quantity 1
Certificate of Analysis no

Additional Information

Additional Information

Manual N/A
Product Specs

Manufacturer

Manufacturer Hanna Instruments

Reviews

Be the first to review this product

How do you rate this product? *

 
1 1 star
2 2 star
3 3 star
4 4 star
5 5 star
Quality
Price
Value